Performance of CsI:Tl Cyrstal with a Spectrum Matching Photomultiplier Tube
This report documents an effort to improve the energy resolution for a thallium doped cesium iodide (CsI:T1) scintillator paired with a spectrum matching photomultiplier tube (PMT). A comparison of the differences in the pulse height spectra from thallium doped (CsI:T1) and sodium doped (CsI:Na) single crystals with PMTs of different spectrum responses was performed. Results show that energy resolution of the detector only improves 0.5% at room temperature when these scintillators are coupled with a spectrum matching PMT. Based on a spectrum matching PMT, the best results for energy resolution are 7.39% and 7.88% for CsI:T1 and CsI:Na scintillators, respectively. The improvement is primarily attributed to the increase of photon statistics from the increase of photons (N) being detected in the spectrum matching PMT. Other factors, such as optical quantum yield and non-proportionality of the CsI:T1 and CsI:Na crystals, that can affect the energy resolution were also studied and reported. The results indicate that although the use of a spectrum matching PMT enhances the photon statistics, it also exacerbates the nonproportionality response. Consequently, a promised improvement on energy resolution due solely to photon statistics was not fully realized.