Publications Details
Quantitative Analysis with Heavy Ion E-TOF ERD
Heavy ion TOF ERD combined with energy detection (E-TOF-ERD) is a powerful analytical technique taking advantage of the following facts: the scattering cross section is usually very high ({approximately}10{sup {minus}21} cm{sup 2}/sr) compared to regular He RBS ({approximately}10{sup {minus}25} cm{sup 2}/sr), contrary to what happens with the energy resolution in ordinary surface solid barrier detectors, time resolution is almost independent of the atomic mass of the detected element, and the detection in coincidence of time and energy signals allows for the mass separation of overlapping signals with the same energy (or time of flight). Measurements on several oxides have been performed with the E-TOF-ERD set up at Sandia National Laboratories using an incident beam of 10-15 MeV Au. The information on the composition of the sample is obtained from the time domain spectrum, which is converted to energy domain, and then, using existing software codes, the analysis is performed. During the quantification of the results, they have found problems related to the interaction of the beam with the sample and to the tabulated values of the stopping powers for heavy ions.