Publications Details

Publications / Journal Article

A comparison of the neutron detection efficiency and response characteristics of two pixelated PSD-capable organic scintillator detectors with different photo-detection readout methods

Marleau, Peter M.; Sweany, Melinda; Balajthy, Jon A.

We characterize the performance of two pixelated neutron detectors: a PMT-based array that utilizes Anger logic for pixel identification and a SiPM-based array that employs individual pixel readout. The SiPM-based array offers improved performance over the previously developed PMT-based detector both in terms of uniformity and neutron detection efficiency. Each detector array uses PSD-capable plastic scintillator as a detection medium. We describe the calibration and neutron efficiency measurement of both detectors using a 137Cs source for energy calibration and a 252Cf source for calibration of the neutron response. We find that the intrinsic neutron detection efficiency of the SiPM-based array is (30.2 ± 0.9)%, which is almost twice that of the PMT-based array, which we measure to be (16.9 ± 0.1)%.