This memo summarizes the results of verification tests on a cone-socket fitting to the ISO 5356-1:2015 standard, annexes B and F. The components were found to meet requirements holding 50N tensile force and more than 35 N-cm of torque for 10 seconds without separating at 35°C with a fresh socket. Humidity control was unavailable; consequently, humidity was significantly lower than the values in the ISO standard. Several repeat tests of the fitting with force, torque, and time excursions give us confidence that the fittings perform suitably to the ISO standard. Force overload tests suggest that the samples will disengage under tension at around 69 N, while torque overload tests under a force of 50 N will fail at 79 N-cm of torque. If repeated tests are performed on the same component, the socket can deform and lead to the connection failing the standard requirements. However, such failures only occur with repeated testing, and the gripping of such units can be modified so that they pass the standard. This observation demonstrates that socket gripping is responsible for those failures and would not be a problem in application.