Publications Details

Publications / Journal Article

High-energy radiographic imaging performance of LYSO

Webb, Timothy J.

A comprehensive comparison of the dominant sources of radiation-induced blur for radiographic imaging system performance is made. End-point energies of 6, 10, 15, and 20 MeV bremsstrahlung photon radiation produced at the Los Alamos National Laboratory Microtron facility were used to examine the performance of large-panel cerium-doped lutetium yttrium silicon oxide (LYSO:Ce) scintillators 3, 5 and 10 mm thick. The system resolution was measured and compared between the various end-point energies and scintillator thicknesses. Contrary to expectations, it is found that there was only a minor dependence of system resolution on scintillator thickness or beam end-point energy. This indicates that increased scintillator thickness does not have a dramatic effect on system performance. The data are then compared to Geant4 simulations to assess contributions to the system performance through the examination of modulation transfer functions. It was determined that the low-frequency response of the system is dominated by the radiation-induced signal, while the higher-frequency response of the system is dominated by the optical imaging of the scintillation emission.