Data Set Details

Data Sets / Figures, plots, charts, and diagrams

2025 Annual INMM Graph and Tables for High Purity Germanium Detector Normalization Presentation

Total Size: 102 KB

The data set includes gamma spectroscopy peak data for measurements taken with two different high purity germanium detectors using a mixed nuclide source and a U-235 fuel rod. There are a total of 5 specific energy peaks that were analyzed for the mixed nuclide source stemming from Am-241, Cd-109, Cs-137, and Co-60. There are a total of 3 specific energy peaks that were analyzed for the U-235 fuel rod. The data set includes the calculations and results from using a linear  correction factor, absolute efficiency curve, and relative efficiency curve to compare the net peaks counts from two different detectors.

Contributors

Drumm, Natalie; Donart, Megan

Data Files

Transfer data using Globus

Source

10.2172/3403493

Fields

There are 9 total tabs included in the data set. The first tab is a combination of the graphs and data set overviews showcasing the final results from comparing the data of the two detectors using a linear correction factor, absolute efficiency curve, and relative efficiency curve. The remaining 8 tabs all relate to the individual data associated with the energy peaks that were used in the calculations.

Sample Data Set

Relative Efficiency

Centroid (keV)
Canberra Peak (CPS)
ORTEC Peak (CPS)
Relative Efficiency (ORTEC/Canberra)

Peak 1
59.54
505.98
55.59
0.1099

Peak 2
88.03
15.68
5.59
0.3567

Peak 3
661.66
90.27
69.92
0.7746

Peak 4
1173.23
44.04
36.29
0.8241

Peak 5
1332.49
38.84
32.84
0.8455

% Difference of Canberra/Ortec

Centroid (keV)
Net Area (Counts)
Peaks (CPS)
FWHM (keV)
RIO Counts

Peak 1
0.203%
89.013%
89.013%
70.057%
86.748%

Peak 2
0.023%
64.327%
64.326%
65.201%
24.989%

Peak 3
0.004%
22.541%
22.541%
32.103%
21.135%

Peak 4
0.025%
17.590%
17.591%
22.782%
17.397%

Peak 5
0.006%
15.450%
15.450%
23.404%
16.796%

Peak 6
0.100%
15.040%
15.042%
49.197%
86.902%

Peak 7
0.000%
32.613%
32.615%
48.893%
52.340%

Peak 8
0.034%
31.831%
31.830%
44.274%
59.565%

 

Absolute Efficiency

Nuclide
Centroid (keV)
Half-Life (days)
Intial Activity (cps)
Current Activity (cps)
Aegis Efficiency
ORTEC Efficiency
% Difference

Peak 1
Am-241
59.54
158000
4.02E+03
3.97E+03
12.74%
1.40%
89.01%

Peak 2
Cd-109
88.03
461.4
5.69E+03
1.23E+02
12.72%
4.54%
64.33%

Peak 3
Cs-137
661.66
10990
3.90E+03
3.32E+03
2.72%
2.11%
22.54%

Peak 4
Co-60
1173.23
1925
7.33E+03
2.93E+03
1.51%
1.24%
17.59%

Peak 5
Co-60
1332.49
1925
7.34E+03
2.93E+03
1.33%
1.12%
15.45%

Supplemental Information

The data is limited in that only 30 measurements were taken for each detector at each peak associated with the mixed nuclide source peaks. Only 20 measurements were taken for each detector at each peak associated with the U-235 fuel rod. Overall the data sets are fairly small and the data was primarily used to check for viability of which of the three options would work best to compare the detectors. This allowed the group to focus on the best option with additional data going forward.

Software Needed

Microsoft Excel

Coverage

Spatial Coverage

Albuquerque, NM

Temporal Coverage

gamma spectroscopy measurements were conducted between June 2021 through July 2021

Top