Overview
Ion Mobility
Spectrometry (IMS) is an attractive
technology for developing a miniaturized
chemical sensor. An IMS sensor system
has the advantage of operating at
atmospheric pressure and is able
to detect trace quantities of materials.
Current IMS drift tube designs require
complex assemblies of electrodes
and insulators coupled with electronics
to produce the uniform electric
fields desired in the IMS drift
tube. This results in systems that
are not suited for mass production
because they are labor intensive
and expensive to produce. We have demonstrated a miniaturized, simple to
assemble, and inexpensive IMS drift tube constructed from "rolled", low-temperature co-fired ceramics (LTCC) with integral potential resistors. The new LTCC design eliminates as many as 150 individual parts from our current "stacked" design
while producing a more chemically inert internal surface.
Applications
- Explosive detection.
- Biological material detection.
- Chemical weapon and drug detection.
Technical Approach

Figure 1: Photograph of LTCC drift tube showing Macor® end pieces, integral
resistors, location of 241Am source, integral heater connections, location of target, and location of apertures and control grids.
Ion Mobility Spectrometry (IMS) is an emerging technique that holds promise for a variety of analytical applications. IMS is essentially an ion filter that sorts the mobility (k) of an ionized molecule; mobility is a measure of how fast a molecule will travel under the influence of an electric field in a given atmosphere. Ion drift time (td) is measured across a known tube length (L) with a known electric field (E) and the mobility is given by the following equation:
L
k = ——
tdE
IMS tubes can operate in either a positive or a negative ion mode depending on the target species. An IMS requires an ionization source to function. This can either be a radioactive source or other means such as a corona discharge source. Our tubes employ an 241Am similar to a smoke detector source as the ionization source.
Low Temperature Co-fire Ceramic (LTCC) Drift Tube
IMS drift tube designs for handheld instruments are constrained by several key
areas of concern. First, the tubes must be small and lightweight enough for the
application. Second, the sensitivity and resolution must be adequate for the
application, and finally, the cost of the manufacture and assembly of the tube
must be kept low enough to be practical. Low temperature co-fire ceramic technology
holds the promise of inexpensive automated drift tube production with minimal
part counts and electrical connections to the control and detection circuitry1.
LTCC allows simple screen-printing operations for the production of the electrodes,
and on-board surface mount resistors fabricated directly on the tube to reduce
the number of electrical connections required without sacrificing the performance
of numerous electrodes (80 electrodes) with small (0.75 mm) spacing. In addition,
integral heating elements are incorporated into the structure rather than applied
to the outside as in the "stacked" design.

Figure 2: Typical data illustrating the operation of an IMS detecting explosives (RDX). Initially the position of the ion peak is around 6ms but moves to the right (10ms) when RDX is injected into the tube. The sensor detects a peak for both the RDX monomer and the RDX dimmer as illustrated. Horizontal axis is the drift time and the vertical axis is the elapsed time.
Most of the processing of the LTCC occurs during its "green" unfired state during which it is still pliable. Thus, the processing can be done to flat sheets of material and the tube is then "rolled" onto
a mandrel of the desired inner diameter and fired. After firing, the resistors
are laser trimmed to within 1% of the desired values and the tubes are dimensionally
trimmed. IMS control structures are constructed using LIGA (LIGA is a German
acronym that stands for lithography, electroforming, and molding) processing
and are inserted into the tubes and form the gate structures, apertures, and
targets of the tube.
Results
MS data similar to that shown in Figure 2 can be analyzed manually or using digital signal processing techniques to determine both the species and the concentration of a target chemical. Figure 2 is a contour plot of the output with the horizontal axis representing the drift time through the IMS tube and the vertical axis representing the elapsed time of the experiment. This particular data is for a 100 ng vapor inhale of RDX and shows both dimer and monomer peaks in the response. IMS has also been demonstrated in a variety of other applications from chemical agent detection to illegal drugs and health applications.
|